Technical specifications – HT instruments HT9021 User Manual

Page 18

Advertising
background image

HT9021

EN - 17

6. TECHNICAL SPECIFICATIONS

6.1. CHARACTERISTICS
Accuracy is calculated as [% rdg + (number of dgt) x resolution]. It is referred to the
following reference conditions: 18°C

÷ 28°C (65°F ÷ 83°F), < 75%RH


DC Voltage (Autorange)

Range Resolution Accuracy

Input

impedance

Overload

protection

600.0mV 0.01mV

±(1.0%rdg + 3dgt)

10M

1000VDC/ACrms

6.000V 0.001V
60.00V 0.01V
600.0V 0.1V

1000V 1V

The meter emits a “bip-bip” sound with VDC >1000V


AC Voltage TRMS (Autorange)

Range Resolution Accuracy

Input

impedance

Bandwidh

Overload

protection

6.000V 0.001V

±(1.0%rdg+4dgt)

(50

÷ 60Hz)

±(3.5%rdg+5dgt)

(61

÷ 400Hz)

10M

50-400Hz 1000VDC/ACrms

60.00V 0.01V
600.0V 0.1V

1000V 1V

Integrated sensor for AC voltage detection: LED turn on for phase-earth voltage > 100V, 50/60Hz
The meter emits a “bip-bip” sound on 1000V range for VAC>770V


Resistance and Continuity test (Autorange)

Range Resolution Accuracy

Buzzer

Overload

protection

600.0

Ω 0.1Ω

±(1.0%rdg+5dgt)

≤ 60

600VDC/ACrms

6.000k

0.001k

60.00k

0.01k

600.0k

0.1k

6.000M

0.001M

60.00M

0.01M

±(2.0%rdg.+10dgt)

Test current of continuity test: < 0.35mA


DC Current

Range Resolution Accuracy

(*)

Overload

protection

60.00A 0.01A

±(2.2%rdg.+10dgt)

1000AACrms

600.0A 0.1A ±(2.0%rdg+8dgt)

1000 1A

(*) Referred to cable inside to the center of clamp jaws. Position sensitivity:

±2.0%rdg


AC TRMS Current

Range Resolution Accuracy

(*) Bandwidh Overload

protection

60.00A 0.01A

±(2.2%rdg+12dgt)

50

÷ 60Hz

1000AACrms

600.0A 0.1A ±(2.2%rdg+8dgt)

1000A 1A

60.00A 0.01A

±(3.5% rdg+12dgt)

61

÷ 400Hz

600.0A 0.1A ±(3.5%rdg+8dgt)

1000A 1A

PEAK function: response time <10ms ; Accuracy PEAK function:

±(5%rdg+10dgt)

(*) Referred to cable inside to the center of clamp jaws. Position sensitivity:

±2.0%rdg