8 technical specifications – HT instruments THT70 User Manual

Page 52

Advertising
background image

THT70

EN - 50

8 TECHNICAL

SPECIFICATIONS

Range Sensitivity Accuracy

(*)

Resolution Image

frequency

-20°C

 400°C <0.06°C@30°C 2%reading or 2°C 384x288pxl 50Hz

(*) With standard lens. For optional lens the accuracy is

4%reading or

4°C choosing the higher value

General specifications
Type of IR sensor

UFPA (384x288pxl, 25

m)

Spectrum response:

8

14m

Visual range (FOV) / Lens:

24.6° x 18.6° / 22mm

IFOV (@ 1m):

1.14mrad

Focusing: manual
Minimum focus distance:

0.3m

Temperature readings:

°C, °F, °K

Available colour palettes:

8 (Standard) + 10 (Custom)

Laser pointer:

Class 2 in compliance with IEC 60825-1 (THT60)

Electronic zoom:

x1, x20

Image rotation:

 360° in steps of 1°

Emissivity correction:

0.01 ÷ 1.00 in steps of 0.01

Measuring modes:

Automatic / Manual / Histogram

Measuring functions:

corrections according to env. temperature, reflected
temperature, distance, relative humidity, offset.

Advanced analyses:

Spots (3), Lines (2), Areas (3), Isotherm

Built-in photo camera:

CMOS sensor 640x480pxl, FOV 62.3°

Fusion PiP Function:

present (IR in Visual & Visual in IR)

Video output:

PAL/NTSC

IR video recording:

by means of micro SD card (MPEG4 format)

Voice annotation:

up to 60s / image

Text annotation:

with virtual keyboard

Memory:

Micro SD card 4GB (>1000 JPG), Max 32GB

PC interface:

USB 2.0

Power supply
Internal supply:

rechargeable Li-ION battery, 7.4V 2700mAh

External supply:

adapter 100-240VAC (50/60Hz)/12VDC

Battery duration:

4.5 hours of continuous use

Display
Characteristics:

Colour TFT LCD 3.5” with capacitive touch-screen

Mechanical characteristics
Size (L x W x H):

243x103x160mm (10x4x6in)

Weight (battery included):

0.92kg (2lv)

8.1 ENVIRONMENT
8.1.1 Environmental conditions for use
Operating temperature:

-20°C ÷ 50°C (-4°F ÷ 122°F)

Storage temperature:

-40°C ÷ 70°C (-40°F ÷ 158°F)

Allowable relative humidity:

10 ÷ 90%RH

Mechanical protection:

IP65 in compliance with IEC 529

Fall test:

2m

Shock:

25G in compliance with IEC60068-2-29

Vibrations:

2G in compliance with IEC60068-2-6

This instrument complies with Directive EMC 2004/108/EC

Advertising