Barotroll instrument – In-Situ Level TROLL 400/500/700/700H Operators Manual User Manual

Page 12

Advertising
background image

1-800-446-7488

12

www.in-situ.com

BaroTROLL Instrument

The titanium BaroTROLL measures and logs barometric pressure and temperature.
Use the BaroTROLL in conjunction with non-vented In-Situ Instruments.

Win-Situ BaroMerge Software simplifies post-correction of water level data by
automatically subtracting barometric readings from data collected by a non-vented
instrument to compensate for changes in pressure due to barometric fluctuations.

General

BaroTROLL

Temperature ranges

1

Operational: -20-80° C (-4-176° F)
Storage: -40-80° C (-40-176° F)
Calibrated: -5-50° C (23-122° F)

Diameter

1.83 cm (0.72 in.)

Length

21.6 cm (8.5 in.)

Weight

197 g (0.43 lb)

Materials

Titanium body; Delrin nose cone

Output options

Modbus/RS485, SDI-12, 4-20 mA

Battery type & life

2

3.6 V lithium; 10 years or 2M readings

External power

8-36 VDC

Memory
Data records

3

Data logs

1.0 MB
65,000
2

Log types

Linear

Fastest logging rate & Modbus rate

1 per minute

Fastest SDI-12 & 4-20 mA output rate 1 per second

Real-time clock

Accurate to 1 second/24-hr period

Sensor Type/Material

Piezoresistive; titanium

Calibrated Range
(Usable Range)

30 psia (usable up to 16.5 psi, 1.14 bar)

Burst pressure

Vacuum/over-pressure above 16.5 psi damages sensor

Accuracy @ 15° C

4

±0.1% FS

Accuracy (FS)

5

±0.2% FS

Resolution

±0.005% FS or better

Units of measure

Pressure: psi, kPa, bar, mbar, mmHg, inHg, cmH

2

O, inH

2

O

Temperature Sensor

Accuracy & resolution

±0.1° C; 0.01° C or better

Units of measure

Celsius or Fahrenheit

Warranty

3 years
Up to 5-year extended warranties are available for all instruments—call for details

Footnotes

See See page 8.

Advertising