3.2 Instrument Set-Up

The following section details the set-up modes available to the user.
These are: Cell constant, Coefficient, Reference Temperature, TDS EC Ratio, Temperature units
and Baud Rate. These can be set in sequence as detailed in this section or, by entering the SET
UP mode and using the STO key to select a specific parameter at any time.

To exit the set up menus at any time press the ESC key. This will return the instrument to the
MODE menu. Any parameters not saved will remain as defaults or previous setting.

3.2.1 Cell Constant (K) entry

Select SETUP mode on the display using the Left arrow key.

The secondary display will show CELL K and will then scroll CELL CONSTANT after 10 seconds.

Symbol will flash while adjustment is being made.

The first digit of the displayed value will be flashing. Adjustment of the flashing digit can be made
using the Up/Down arrow keys.

The Left/Right arrow keys are used to select the next adjusting digit which will flash when selected.

Once the full reading has been adjusted to the required value, press the STO key. The symbol will
disappear and the secondary display will momentarily show SAVED. The instrument display will
update and then show the Coefficient set up menu.

7

451 050/REV A/03-03