Jenway 6320D User Manual

Page 13

Advertising
background image

13

1. Lamp access panel

Allows access to lamp when replacement is necessary.

NOTE: The lamp access panel and all ventilation slots must NOT be covered or

obstructed at any time.

2. Power switch

On/off switch for the unit

3. Power in socket

IEC type connection socket for mains cable

4. Output sockets

Analogue output

5. RS232 serial port

Connection to a PC

2.6

SAMPLING

Model 6300 is supplied with a 10mm cuvette holder. Other accessories can be fitted (refer to
section 7).

Model 6320D is supplied with a dual 10mm cuvette holder with a 12.7mm test tube holder. The
domed lid on this model enables tall test tubes (up to 100mm in height) to be used.

WARNING: The optical characteristics and light path of the cuvettes and test tubes used in the

dual purpose sample holder vary so cannot be interchanged. Calibrate on one and use only the
same type of sample holder for each analysis batch.