Common trouble shooting – KANOMAX 6710 TABmaster Capture Hood User Manual User Manual

Page 31

Advertising
background image

24

7. Common Trouble Shooting

Symptom

Possible causes / troubleshooting

No display

Check the positive/negative connections of the batteries.
→Verify the batteries are inserted correctly.
Low battery charge.
→Replace or recharge the batteries.

Main structure cannot be adjusted

Low battery charge.

→Replace or recharge the batteries.

A test is running.
→Stop the test, adjust the incline settings of the instrument
again.
Check the interface.
→Return to the main interface.

“DATA Error” after test in
B. Press mode

Flap closed pressure was less than flap open pressure, or flap
open versus flap closed pressure ratio exceeds allowed limits.
→Test again.

Test data was not saved after
testing

The Cycle is full in the current ID.
→Add a new ID.

A new ID cannot be added

The current ID is already a new ID.
→Use the current ID.
Check the ID No.
→If the ID No is 30, delete some IDs.

Print appears garbled

Baud rate does not match.
→Set Baud again.

Display “Connect Error”

Check the connection between the instrument and the cable.
→Connect the cable again and restart.


The test does not begin after
pressing “START”

Check the connection between the instrument and the cable.
→Check the cable, and connect again.
Low battery charge.
→Replace or recharge the batteries.
The instrument is being adjusted.
→Stop adjusting the instrument.
Check the back pressure plate.
→Close the Back pressure plate.

Test exception when testing

Check the connection between the instrument and the cable.

→Check the cable, and connect again.

Abnormal operation of the
solenoid valve when testing

Check the connection between the instrument and the cable.

→Check the cable, and connect again.

Motor movement disorders when
testing

Check the connection between the instrument and the cable.

→Check the cable, and connect again.

Advertising