4 changing the calibration password, Table 4-1. factory default calibration passwords, Changing the calibration password -3 – KEPCO ABC-DM SERIES User Manual

Page 61: Factory default calibration passwords -3, 4 to c

Advertising
background image

ABC 040104

4-3

4. LCD will flash

Iout F.S.adj. **WAIT**

then

Iout F.S. adj < >

. Monitor DVM and

use

or

keys to increase or decrease output voltage as necessary to get DVM reading

as close as possible to value listed below (without exceeding value listed). If the

"

and

#

keys do not provide the required adjustment range, press the 1 (increase) or 3 (decrease)
key for additional range.

ABC 10-10DM:

1.000V

ABC 15-7DM:

0.700V

ABC 25-4DM:

1.000V

ABC 36-3DM:

0.750V

ABC 60-2DM:

2.000V

ABC 125-1DM:

1.000V

5. Press ENTER when complete.

4.4

CHANGING THE CALIBRATION PASSWORD

The 4-digit password is required for to enter calibration mode. The factory default passwords
are listed in Table 4-1.

1. If the unit is not in command entry status , press CLEAR or RESET to put the unit in com-

mand entry status.

2. Press MENU key until LCD reads

Change Passw. OldPass

.

3. Enter the old password and press ENTER. Then enter the new 4-digit Password and press

ENTER.

4. The new password is accepted and the LCD shows the next Menu screen

GoTo Previous

CALIB? 1 Yes

.

5. Press CLEAR to exit the Menu, press 1 to restore the previous calibration values (see PAR.

4.5), or MENU for additional menu screens.

TABLE 4-1. FACTORY DEFAULT CALIBRATION PASSWORDS

MODEL

PASSWORD

ABC 10-10DM

1010

ABC 15-7DM

1507

ABC 25-4DM

2504

ABC 36-3DM

3603

ABC 60-2DM

6002

ABC 125-1DM

1251