3.6 Preparing and connecting the signal and field current cables (only

TIDALFLUX)

3.6.1 Cable lengths

Interface cable

Interface cable

Interface cable

Interface cable: maximum length is 600 m / 1968 ft.

Type B (BTS) signal cable

Type B (BTS) signal cable

Type B (BTS) signal cable

Type B (BTS) signal cable: maximum length is 600 m / 1968 ft.

Type A (DS) signal cable

Type A (DS) signal cable

Type A (DS) signal cable

Type A (DS) signal cable: maximum length depends on the conductivity of the fluid:

Field current cable

Field current cable

Field current cable

Field current cable: The cross section of the cable determines the maximum length:

DANGER!
Cables may only be connected when the power is switched off.

DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.

CAUTION!
The maximum allowed distance between the flow sensor and the converter is determined by the
shortest cable length.

Electrical conductivity

Maximum length

[µS/cm]

[m]

[ft]

50

120

394

100

200

656

200

400

1312

≥400

600

1968

Cross section

Maximum length

[mm

2

]

[AWG]

[m]

[ft]

2 x 0.75

2 x 18

150

492

2 x 1.5

2 x 14

300

984

2 x 2.5

2 x 12

600

1968