SERVICE
7
55
UFM 530 HT
www.krohne.com
12/2013 - 4002393201- HB UFM 530 HT R02 en
The reassembling takes place in the reverse order. Please regard the following items:
• Sensor specific data must be entered in the converter.
• All application specific settings must be re-entered after replacement of the electronic unit.
The supplied report on settings contains the standard factory setting.
• It is recommended to check the zero point and eventually perform a new zero measurement.
Figure 7-3: Disconnecting display cable
1 Pull the display towards you
2 Pull out the connector from the electronics.
Figure 7-4: Removing electronics unit
1 Loosen the three screws as indicated.
2 Take out the electronics unit.