Service – KROHNE UFM 530 HT EN User Manual

Page 55

Advertising
background image

SERVICE

7

55

UFM 530 HT

www.krohne.com

12/2013 - 4002393201- HB UFM 530 HT R02 en

The reassembling takes place in the reverse order. Please regard the following items:

• Sensor specific data must be entered in the converter.
• All application specific settings must be re-entered after replacement of the electronic unit.

The supplied report on settings contains the standard factory setting.

• It is recommended to check the zero point and eventually perform a new zero measurement.

Figure 7-3: Disconnecting display cable

1 Pull the display towards you
2 Pull out the connector from the electronics.

Figure 7-4: Removing electronics unit

1 Loosen the three screws as indicated.
2 Take out the electronics unit.