DESCRIPTION OF HART INTERFACE

30

SMARTSENS ORP 1590

www.krohne.com

11/2013 - 4002547102 - MA SMARTSENS ORP 1590 R02 en

7.4 Inputs/outputs and HART

®

dynamic variables and device variables

PV = Primary Variable; SV = Secondary Variable; TV = Tertiary Variable

Code = device variable code

Device variables

The HART

®

dynamic variable PV is always connected to the HART

®

current output which is

assigned to the ORP value.

7.5 Field Communicator 475 (FC 475)

The Field Communicator is a hand terminal from Emerson Process Management that is
designed to configure HART

®

and Foundation Fieldbus devices. Device Descriptions (DDs) are

used to integrate different devices into the Field Communicator.

7.5.1 Installation

The HART

®

Device Description for the sensor must be installed on the Field Communicator.

Otherwise only the functions of a generic DD are available to the user and the entire device
control is not possible. A "Field Communicator Easy Upgrade Programming Utility" is required to
install the DDs on the Field Communicator.

The Field Communicator must be equipped with a system card with "Easy Upgrade Option". For
details consult the Field Communicator User’s Manual.

HART

®

dynamic variable

PV

SV

TV

ORP

Electrode voltage

Temperature

HART

®

device variable

Code

Type

ORP value

0

linear

Electrode voltage

1

linear

Temperature

2

linear