Description of hart interface, 8 basic-dd menu tree (details for settings) – KROHNE SMARTSENS ORP 1590 EN User Manual

Page 32

Advertising
background image

7

DESCRIPTION OF HART INTERFACE

32

SMARTSENS ORP 1590

www.krohne.com

11/2013 - 4002547102 - MA SMARTSENS ORP 1590 R02 en

7.8 Basic-DD menu tree (details for settings)

1 Measuring values

1 ORP value

Display of the measured ORP value on process control system, HART

®

handheld or display.

2 Temperature

Display of the measured temperature value in °C / °F on process control
system, HART

®

handheld or display.

3 Electrode voltage

Display of the measured electrode voltage in mV on process control system,
HART

®

handheld or display.

4 Loop current

Display of the measured loop current in mA on process control system,
HART

®

handheld or display.

5 Error

Display of status messages and diagnostic information. For further

information refer to

Status messages and diagnostic information on page 19.

6 Device state

Display of status information icon for sensor according to NAMUR NE 107.

2 Quick setup

1 TAG

Set the TAG number for the measuring loop.

2 Manual hold

Activate or deactivate the manual hold function. Select between yes or no.

3 I/O

Meas. value at 4 mA

Set measuring value at 4 mA;

Default setting: -1500 mV

Meas. value at 20 mA

Set measuring value at 20 mA;

Default setting: +1500 mV

Time constant

Set time constant (1…60 seconds);

Default setting: 1 second

4 Calibration

Start calibration procedure. For more information refer to

Calibration on

page 17.

5 Input user password

Set password

3 Logbooks

1 Calibration logbook

Logbook for the last 20 calibration cycles

2 Error logbook

Display errors

.book Page 32 Wednesday, October 30, 2013 11:26 AM

Advertising