8.6 Measuring accuracy (only TIDALFLUX)

The measuring accuracy for partly filled pipes and completely filled pipes are different. In these
graphs it is assumed that the velocity at full scale value is at least 1 m/s (is also the standard
value for calibration, since it will result in the most accurate measurements).

Maximum measuring error

Related to volume flow (mv = measured value,

FS = Full Scale)

These values are related to the pulse / frequency output

The additional typical measuring deviation for the current output is
±

10 μA

Partly filled:

Partly filled:

Partly filled:

Partly filled:

v ≥ 1 m/s / 3.3 ft/s at Full Scale: ≤ 1% of FS

Fully filled:

Fully filled:

Fully filled:

Fully filled:

v ≥ 1 m/s / 3.3 ft/s: ≤ 1% of mv

v < 1 m/s / 3.3 ft/s: ≤ 0.5% of mv + 5 mm/s / 0.2 inch/s

Minimum level: 10% of inner diameter

Fully filled pipes

Figure 8-1: Maximum measuring error of measured value.