KROHNE IFC 090 Converter EN User Manual

Page 48

Advertising
background image

6/3

6.4 Pulsating flow

Application
downstream of positive-displacement pumps (reciprocating or diaphragm pumps) without
pulsation dampener

Operator control of the signal converter for the new settings
IFC 090 B (basic version)

see Sect. 6.2

IFC 090 D (display version) see Sect. 4 and 5

To change settings
Fct. 3.02 FIELD FREQ. (change magnetic field frequency)

– Stroke frequency less than 80 strokes/min (at max. pump lift): do not change setting.
– Stroke frequency 80 - 200 strokes/min (at max. pump lift): change setting to 1/2,

only practical with IFM 5080 K and IFS 5000 F (DN 2.5-100 /

1

/

10

”-4”)

and IFM 4080 K and F IFS 4000 F (DN 10, 15, 50-100 /

1

/

10

”,

1

/

2

”, 2”-4”),

Please consult factory where other types and meter sizes are concerned.

– Please note: given stroke frequencies close to the tripping point of 80 strokes/min, additional

measuring errors of approx. ± 0.5% of the measured value may occasionally occur.

• Fct. 3.06 APPLICAT. (adjust overload point of the A/D converter to the application)

Change setting to “PULSATING”.

• Fct. 1.04 DISP. FLOW (change display presentation of flow)

Change setting to “BARGRAPH” to allow better assessment of display unsteadiness.

• Fct. 1.02 TIMECONST. (change time constant)

– Set to “ALL” and time (t) in seconds.

– Recommended:

t [s] =

– Example:

min. number of strokes in operation = 50 strokes/minute

t [s] =

= 20 s

With this setting, the residual ripple of the display will amount to
approx. ± 2% of the measured value. Doubling the time constant
will reduce the residual ripple by a factor of 2.

1000

min. strokes / min

1000

50 / min

6.5 Rapid changes in flowrate

Application
in conjunction with batching processes, fast-response control loops, etc.

Operator control of the signal converter for the new settings
IFC 090 B (basic version)

see Sect. 6.2

IFC 090 D (display version) see Sect. 4 and 5

To change settings
• Fct. 1.02 TIMECONST. (change time constant)

Setting to “ONLY I” and set time to 0.2 s.

• Dynamic response with meter sizes DN 2.5-300 /

1

/

10

”-12”

Dead time:

approx. 0.06 s at 50 Hz line frequency
approx. 0.05 s at 60 Hz line frequency

Time constant: set as above, current output (mA) additionally plus 0.1 s

• Reducing the dead time by a factor of 3 (possible by changing the magnetic field frequency)

Fct. 3.02 FLOW METER, subfunction “FIELD FREQ.”, change to “1/2”,
only practical with IFM 5080 K and IFS 5000 F (DN 2.5-100 /

1

/

10

”-4”)

and IFM 4080 K and IFS 4000 F (DN 10, 15, 50-100 /

1

/

10

”,

1

/

2

”, 2”-4”)

Please consult factory where other types and meter sizes are concerned.