KROHNE CORIMASS E Series EN User Manual

Page 44

Advertising
background image

44

Fct. No.

Text

Description and settings

3.9.0

TUBE PARAMS

Submenu 3.9.0 Transducer calibration and

compensation parameters*

3.9.1

Fgw

CF1

Drive frequency - water: from Calibration Certificate

3.9.2

Fcw

CF2

Coriolis frequency - water: from Calibration

Certificate

3.9.3

Fgl

CF3

Drive frequency - air: from Calibration Certificate

3.9.4

Fcl

CF4

Coriolis frequency - air: from Calibration Certificate

3.9.5

GK

CF5

Transducer mass flow calibration constant: from

Calibration Certificate

3.9.6

LIN

CF6

Linearity adjustment: from Calibration Certificate

3.9.7

Tcl

CF7

Mass flow temperature compensation: from

Calibration Certificate

3.9.8

Tc0

CF8

Mass flow temperature compensation at zero flow:

from

Calibration Certificate

3.9.9

TcD

CF9

Density temperature compensation: from Calibration

Certificate

3.9.10

D.REF.WATER

Use

key to select between two modes, then press

to

select

submenu and then

for choice of parameters

* MEAS.VALUE

CALIB. NO

CALIB. YES

* SET VALUE

freq Hz

temp

°

C (

°

F)

density g/cm

3

3.9.11

D.REF.AIR

Use

key to select between two modes, then press

to

select

submenu and then

for choice of parameters

* MEAS.VALUE

CALIB. NO

CALIB. YES

* SET VALUE

freq Hz

temp

°

C (

°

F)

density g/cm

3

* Many, but not necessarily all, compensation parameters are shown on instrument data
plate.

These menus, except 3.9.10 and 3.9.11, are protected by the Code 4 password; see
Fct. 3.8.8.