Example 2 :
Measured value

= temperature

Max temperature

= 75°C

Min temperature

= -25°C

Max frequency

= 1000 Hz

Temperature

Frequency

< - 25°C

0°C

20°C

75°C

> 95°C

0 Hz

250 Hz

450 Hz

1000 Hz

1300 Hz

To set up example 1, proceed as follows :

Begin from measuring mode.

The frequency output will allow flows to be registered up to 1.3 times the maximum value.
(Note : for mass flow and volume flow, all flows are assumed positive).
The absolute maximum output frequency is 1300 Hz so the Max. Value allowed for Fct. 3.4.2. is
1000 Hz allowing for the 1.3 x over-range. See Figure 35 for the frequency output characteristics
of examples 1 and 2.

NOTE:
The frequency output has a duty cycle of 50% for a frequency range > 1Hz. For a frequency
range < 1Hz, the duty cycle is not 50%.

Key

Display
line line

2

→↑↑

Fct. (3).0.

INSTALL.

→↑↑↑

Fct. 3.(4).0.

PULS.OUTP. P

Fct. 3.4.(1).

FUNCTION P

(TOTAL MASS)

(MASS FLOW)

Fct. 3.4.(1).

FUNCTION P

Fct. 3.4.(2).

PULSE/TIME

(0)1000

MAX Hz

Current max. frequency 1000 Hz.

9x

(0)000

MAX Hz

0(0)00

MAX Hz

5x

0(5)00

MAX Hz

Fct. 3.4.(2).

PULSE/TIME

Fct. 3.4.(3).

MIN. FLOW

Input min. flow of 0 kg/min

↵↑

Fct. 3.4.(4).

MAX. FLOW

Input max. flow of 5 kg/min

Fct. 3.4.(4).

MAX. FLOW