Installation and Operating Instructions OPTIMASS

78

5.3 Menu 3- Configuration Menu

To access this menu, enter programming mode

Display

Key

Line 1

Line 2

Fct. 1

Operation

Fct. 2

Test

Fct. 3

Config.

Fct. 3.1

Basic Param.

Fct. 3.1.1

Low Flow cut off


Low Flow Cut Off Fct. 3.1.1

If the flow mode in (Fct. 3.1.3. is set to Flow +/- then at zero flow small signal fluctuations will average out

to nothing and the totaliser will remain static.

However, if ‘unidirectional’ flow is selected, this process will not work and the totaliser reading will

increment slowly with time. To prevent this, the Low Flow cut off should be set.

The Low Flow cut off is entered as a percentage of the rated nominal flow of the primary head. The cut off

may be set from 0 to 10% in increments of 0.1%. Thus for example a 7000 T25 (nominal flow 34500 kg/h

or 1250 lbs/min) with a low flow cut off of 0.2%, any flow less than 69kg/hr or 2.5 lb/min will indicate zero

on the display.

To set the low flow cut off to 1%:

Display

Key

Line 1

Line 2

From

Fct. 3.1.1

Low Flow cut off

(0)0.5 Percent

(0).5 Percent

(1).5 Percent

1.(5) Percent

Repeat till (0) then

to accept the value.


Time Constant Fct. 3.1.2

Measurements taken from the sensor are digitally filtered to provide stable readings in the event of

fluctuating flows. The degree of filtering also affects the response time of the reading due to rapid

changes in the flow.

Short time constant:

Fast response

Fluctuating reading

Long time constant:

Slow response

Stable reading

The figure below shows the typical response of the measurement for varying time constants and a rapid

change in flow.