Commissioning / operation – KROHNE OPTIWAVE 5200 PROFIBUS User Manual

Page 19

Advertising
background image

COMMISSIONING / OPERATION

4

19

OPTIWAVE 5200 C

www.krohne.com

10/2013 - 4003066901 - MA OPTIWAVE5200C-PA-R01-en-

58

3

Converter ROM Error

Replace the signal converter. For more data, refer to the

handbook.

59

3

Sensor EEPROM Error

Replace the signal converter. For more data, refer to the

handbook.

60

3

Sensor RAM Error

Replace the signal converter. For more data, refer to the

handbook.

61

3

Sensor ROM Error

Replace the signal converter. For more data, refer to the

handbook.

63

3

Oscillator Frequency Failure

Replace the signal converter.For more data, refer to the

handbook.

64

3

Converter Voltage Error

Replace the signal converter. For more data, refer to the

handbook.

65

3

Sensor Voltage Error

Do a check of the power supply at the device terminals. Make sure

that voltage values are in the specified limits in menu item 2.2.2

DIAGNOSTIC (Configuration mode / Supervisor menu). If the

voltage is correct, replace the signal converter. For more data on

how to replace the signal converter, refer to the handbook.

66

3

Internal Communication Error

De-energize the device. Make sure that the signal cable engages

in the terminal and the screw connection is tight. Energize the

device. If the problem continues, replace the signal converter. For

more data, refer to the handbook.

67

3

Temperature Out of Range Error

Measure the ambient temperature. Deenergize the device until

the ambient temperature is back in the given range. If the

temperature does not stay in the correct range, make sure that

there is insulation around the signal converter. If this error

occurs 2 times, replace the device.

68

3

Sensor not compatible

Go to menu 1.1.0 IDENT. in Configuration mode. Record the

version numbers of the device software given in menu items 1.1.2,

1.1.3 and 1.1.4. Give this data to the supplier.

69

3

Sensor no signal

Replace the signal converter. For more data, refer to the

handbook.

72

3

Peak Lost Error

Do a check of the device, tank and the process. Reconfigure the

device and record a new empty spectrum. Follow the instructions

in the handbook. If necessary, speak to the supplier.

73

3

Overfill Error

Use a different procedure to measure the level in the tank.

Remove some of the product until the level is below the blocking

distance.

If a viscous product touched the antenna:

remove the device and clean the antenna, or

if the device has a purging option, use the purging system to
clean the antenna.

82

3

Empty spectrum invalid

Record a new empty spectrum. Follow the instructions in the

handbook.

83

3

Signal weak

If this error occurs frequently, it is possible that you must install

the device in a stilling well or use a different antenna type.

84

3

Signal strong

No corrective action is necessary.

85

3

Bad measurement quality

Record the process with the PACTware™ software tool to find the

cause of the problem. If necessary, speak to the supplier.

Bit Number

sup

p

orted

De

sc

ri

ption

Corrective action