KYORITSU 1062 User Manual

Page 13

Advertising
background image

71

AC Voltage Measurement [RMS]

V

KEW1062

AC Coupling, Rms-value detection, Crest factor*: <3

Range

Reso-

lution

Accuracy

Input

Impedance

Maximum

Input

Voltage

10 to

20Hz

20Hz to

1kHz

1k to

10kHz

10k to

20kHz

20k to

50kHz

50k to

100kHz

50mV 0.001mV 2+80

*2

0.4+40

*2

5+40

*2

5.5+40

*2

15+40

*2

11MΩ

<50pF

1000V

rmsAC

1000V

DC

500mV 0.01mV

1+30

*1

0.4+30

*1

1+40

*1

2+70

*2

5+200

*2

5V 0.0001V

50V 0.001V

10MΩ

<50pF

500V

0.01V

1000V*

0.1V

*2

*2

3+30

*2

-

*: Crest factor <1.5 at 1000V range
Accuracy *1: At 5 to 100% of range, *2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1kΩ)
Response time: 1 sec max.

AC Voltage Measurement [MEAN]

V

KEW1062

AC Coupling, MEAN value detection, RMS value calibration (sine wave)

Range

Reso-

lution

Accuracy

Input

Impedance

Maximum

Input

Voltage

10 to 20 Hz

20 to 500 Hz 500Hz to 1 kHz

50mV 0.001mV

4+80

*2

1.5+30

*2

5+30

*2

11MΩ

<50pF

1000V

rmsAC

1000V

DC

500mV 0.01mV

2+30

*1

1+30

*1

3+30

*1

5V 0.0001V

50V 0.001V

10MΩ

<50pF

500V

0.01V

1000V

0.1V

*2

*2

*2

Accuracy *1: At 5 to 100% of range, *2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1kΩ)
Response time: 1 sec max.