Caution – KYORITSU 1062 User Manual

Page 30

Advertising
background image

88

4.3.9 Temperature Measurement (TEMP)

CAUTION

To avoid damage to instrument

Turn off the power to the circuit under test before starting measurement in
order to prevent any excessive voltage from being applied to the instrument.

Note

Optional Temperature probe is required for temperature measurement.
Temperature Probe: Thermocouple Type K

Model: 8405, 8406, 8407, 8408

Check the measurable range of respective probes.

1) Turn the function switch to the “ TEMP ” position.

2) Plug the measuring probe into the input terminals.

3) Contact the measuring probe to the under test and then read the value

when it stabilizes.

Measuring

probe

Red

Black

Note

The default temperature read-out of the Digital Multimeters is in Celsius (°C).
To change it to Fahrenheit (°F), it is necessary to proceed as follows:
Changing the temperature unit setting to Fahrenheit

Displaying “ ºC ” only is configured at factory before shipment.
Carry out the following setting procedure to display “ ºF ”.
While pressing the SELECT, RANGE and HOLD keys simultaneously, turn the function switch
to the “ TEMP ” position. Then, upon pressing the SELECT key, the temperature unit switches
from ºC to ºF.
Once the temperature is displayed in ºF, press the SELECT key to alternately switch the
temperature units between ºF and ºC.

The conversion from Celsius to Fahrenheit is performed using the equation
Fahrenheit temperature = 1.8×Celsius temperature + 32