KYORITSU 3323A User Manual

Page 7

Advertising
background image

― 5 ―

MODEL3323A

* Heavy-line circular arc on the scale indicates the primary effective

measuring ranges. (guaranteed accuracy range)

<Operating error>

Operating error (B) is an error obtained under the nominal
operating conditions, and calculated with the intrinsic error (A),
which is an error of the instrument used, and the error (En) due to
variations. According to IEC61557, the maximum operating error
should be within +/-30%.

B =︱A︱+ 1.15 ×

( E

12

+ E

22

+ E

32

A : Intrinsic error (%)
B : Operating error (%)

1

: Variation due to changing the temperature (%)

2

: Variation due to changing the Battery voltage (%)

3

: Variation due to changing the position (%)

Nominal operating conditions

Ambient temperature : 0

40℃

Relative humidity

: 90% or less

External magnetic field : 400A/m or less
Position

: Horizontal

±90°

Battery voltage

: within "BATTERY. GOOD" range

Nominal voltage
Max. effective scale value
Mid-scale value
Accuracy in primary
effective measuring ranges
Accuracy in primary
effective measuring ranges
Accuracy at 0 and ∞
Accuracy at no-load
voltage
Nominal current
Short-circuit current

25V

10MΩ

0.2MΩ

0.01∼5MΩ

within ±5% of indicated value

Measuring ranges other than above, 0 and ∞.

within ±10% of indicated value

within ±0.7% of scale length

0%∼+20% of nominal voltage

1mA 0%∼+20%

within 1.5mA

50V

10MΩ

0.2MΩ

0.01∼5MΩ

100V

20MΩ

0.5MΩ

0.02∼10MΩ

Advertising