KYORITSU 3323A User Manual

Page 8

Advertising
background image

― 6 ―

<AC voltage range>

<Number of measurement>

Possible number of measurement within the "BATTERY.GOOD"
range.(Measure 5 sec., and take pause for 25 sec.)

Temperature & humidity

: 0℃∼40℃ (RH: 85% or less)

range

(no condensation)

Storage temperature &

: -20℃∼+60℃ (RH: 75% or less)

humidity range

(no condensation)

Response time

: Meter reading comes within accuracy

within 3 sec. after a resistance
corresponding to the mid-scale value
and 0Ω is suddenly applied across
the measuring terminals.

(It may take time when measuring a

capacitive load.)

Insulation resistance

: at least 100MΩ/ DC1000V between

the electrical circuit and the enclosure

Withstand voltage

: AC5, 550V(50/60Hz) for 1 min. between

the electrical circuit and the enclosure

Measured voltage

Accuracy

0∼600V
within ±3% of max scale value (ACV range)
within ±5% of max scale value (Others)

Range

25V/10MΩ

50V/10MΩ

100V/20MΩ

125V/20MΩ
250V/50MΩ

500V/100MΩ

1000V/2000MΩ

0.025MΩ

0.05MΩ

0.1MΩ

0.125MΩ

0.25MΩ

0.5MΩ

1MΩ

at least 4000 times
at least 3500 times
at least 3000 times
at least 2500 times
at least 2000 times
at least 1000 times
at least 300 times

Possible number
of measurement

Resister

for test