12. Wiring check

Proper wirings can be checked at

WAVE Range.

1. Ordinal screen

2. Checking wiring

3. Check completes

Press the

F2

Key.

Wiring check starts.

[Checking status] &

[Proper vector]
are displayed.

Wiring check completes.

In case of NG, Error message

appears. (Press the

ENTER

Key

when OK is displayed.)

  

* Check results may by affected if great power factors exist at the measurement sites.

Criteria of Judgment and cause

Check

Criteria of Judgment

Cause

Frequency

Frequency of V1 is between 42 and

68Hz.

• Voltage clip is firmly connected to the DUT?

• Measuring too high harmonic components?

Voltage input Voltage input is 10% or more of

(Voltage Range x VT).

• Voltage clip is firmly connected to the DUT?

• Voltage test leads are firmly connected to the

Voltage input terminals on the instrument?

Voltage

balance

Voltage input is within ±30º of

reference voltage (V1)

* (not judged by single-phase

wiring)

• Setting against the wiring under test are matched?

• Voltage clip is firmly connected to the DUT?

• Voltage test leads are firmly connected to the

Voltage input terminals on the instrument?

Voltage phase Phase of voltage input is within ±10º

of reference value (proper vector).

• Voltage test leads are properly connected?

(Connected to proper channels?)

Current input Current input is 5% or more of

(Current Range x CT).

• Clamp sensors are firmly connected to the Power

input terminals on the instrument?

• Setting for Current Range is appropriate for input

levels?

Current phase Current input is within ±60º of

reference value (proper vector).

• Arrow mark on a Clamp sensor and the orientation

of flowing current is matched? (Power supply to

Load)

• Clamp sensors are connected properly?