KYORITSU 1710 User Manual

Voltage tester

Advertising
background image

INSTRUCTION MANUAL



VOLTAGE TESTER

KEW 1700 / 1710


KYORITSU ELECTRICAL
INSTRUMENTS WORKS, LTD

.

1. Features

●Designed to meet international safety standards.
IEC61243-3 / IEC61010-1
Measurement Category (CAT.) IV 600V
●Self-Diagnostic test
●AC and DC voltage test up to 690V with LEDs
and LCD ( only KEW1710 ).
●Polarity indication
●Single-pole phase test
●Phase rotation test
●Continuity test
●Auto-power ON / OFF
●Pen light for illuminating measurement points
●Two way clip for adjustable spacing of probes
●Selectable probe Tips 1.6/4mm
●Probe cover protects user and Tips
●IP65 (IEC60529)
●Compact design (Light weight and portable)

2. Safety Warnings
This instrument has been designed, manufactured
and tested according to IEC 61010/61243: Safety
requirements for Electronic Measuring apparatus,
and is supplied having passed rigorous quality
procedures.

This instruction manual contains warnings and
safety rules which have to be observed by the
user to ensure safe operation of the instrument
and to maintain it in safe condition. Therefore,
read through these operating instructions be-
fore using the instrument.

WARNING is reserved for conditions and ac-

tions that are likely to cause serious or fatal injury.

CAUTION is reserved for conditions and ac-

tions that can cause injury or instrument damage.

It is essential that the above instructions are
adhered to. Failure to follow the above
instructions may cause injury, instrument
damage and/or damage to equipment under test.

Symbols used on the instrument

User must refer to the explanations in
the instruction manual.

Instrument with double or reinforced
insulation, Class II insulation

Insulated personnel body protective
equipment up to 690V.

CAT.IV

The source of the low-voltage
installation:
Eg. Electricity meters and primary
overcurrent protection devices.
Estimated transient overvoltage 8kV.

CE

Comply with EMC and Low Voltage
Directive

WARNING

●Never make measurement on a circuit in
which the electrical potential exceeds 690V.
●Do not attempt to make measurement in the
presence of flammable gasses, as the use
of the instrument may cause sparking,
which could lead to an explosion.
●Never attempt to use the instrument if it’s
surface or your hands are wet.
(Do not use in rainfall.)
●Keep your hands and fingers behind the
barriers during measurements.
●Never unlock and open the Battery case
during measurements.
●Verify proper operation on a known source
before use or taking action as a result of the
indication after use.
●Never attempt to make any measurement if
any abnormal conditions, such as a broken
case or exposed metal parts are present on
the Instrument or test probes.
●Do not make any modification to the
instrument.
●Extreme caution when Live circuit LED

flashes or lights on.
●Correct indication of LEDs is only guaran
teed within a temperature range of -10°C up
to

55°C (<85% RH).

3. Instrument layout















1

12/24/50/120/230/400/690V LEDs

for voltage indication

2

Live circuit LED for Single-pole phase and

Double-pole test

3

L/R LEDs for phase rotation test

4

Power LED

5

Polarity indication LEDs for voltage

6

LCD (only KEW1710 )









































1

L1 probe -

2

L2 probe + (Instrument probe)

3

4mm tips

4

Barrier

5

Pen light

6

Pen light switch

7

Battery case

8

Probe clip (two way)

9

Probe protection cover

1

2

3

4

5

1

2

4

4

5

3

6

7

8

9

R

6