Fuse Testing


The calibrator employs fuses to protect the outputs and inputs in the
event the calibrator is improperly connected to external circuitry. It is
necessary to remove the covers in order to access the fuse panel. The
following tests should be performed BEFORE removing the covers to
determine if a fuse is actually blown.












Service Information

130/135

60

Fuse Type/

Location

Fuse

No.

Function

Test

Input/
Output


Top PCB

130/135

F1

mA Input

Connect an ohmmeter across the
input milliamps terminals. The
resistance should be approximately
5 ohms. If it reads open, the fuse is
probably blown. If it reads
significantly higher than 5 ohms, the
input dropping resistor may have
been damaged due to an overload.

Model 130

Model 135

F4

F2

Freq

Output

Check for normal operation on the
Volts output 10V scale. If this is
normal, but there is no output Freq
mode on the frequency terminals,
the fuse may be blown.