Metex 2200R ORP Analyzer/Controller User Manual

Page 20

Advertising
background image

Model 2200R

Page 18

5.12 Alarm Low

5.12.1 To set the ALARM LOW proceed as follows:


a) Press CALL to enter the menu, proceed by pressing CALL until the ALARM

LOW LED is illuminated. With the arrows make the display read the desired
value. Press ENTER. The display will flash until ENTER is pressed again to
confirm

entry.

b) Press RUN to place the instrument on line, or press CALL, for another menu

selection.

NOTE: A low alarm point higher than the high alarm point is an invalid entry, which

is indicated by the LED flashing when you return to RUN. To correct, return
to the menu and reset the alarm points.


5.13 Test

5.13.1 To check the setup of the 2200R, the TEST feature can be used. The test feature

will allow the user to use the arrow keys to change the display. By changing the
display value the analog outputs and relay outputs will react, as if the process was
actually changing. This feature enables the user to simulate the process and test is
the auxiliary devices connected to the relays are functioning properly.

3.5.1

To use the TEST feature proceed as follows:

a)

To enter the menu press CALL, the TEST LED will be illuminated. Use the
arrow keys to change the display. The outputs will react to the changes in
the display, as if the unit was online and the process was actually changing.


b)

Press RUN to place the instrument on line, or press CALL, for another
menu selection.

5.14

Temperature

5.14.1 If your probe is a standard probe it will not be fitted with a temperature sensor and

therefore the 2200R will not monitor temperature. The TEMPERATURE item in the
menu will be skipped.

3.5.1

However if your probe has been ordered with an integral temperature sensor the
temperature of the process can be read at any time by entering the menu and
scrooling to TEMPERATURE. Either

°C or °F will be indicated depending on the

position of the dip switch.


3.5.2

The 0-5Vdc and 0-1mA analog outputs can be dedicated to follow the process
temperature by simply placing dip switch No.8 on S1 in the OFF position. The
temperature span of the output is set in the Utility Menu. Refer to Section 7.5









Advertising