Metex 2200C Conductivity Analyzer/Controller User Manual

Page 3

Advertising
background image

Model 2200C

Page 1

MODEL 2200C CONDUCTIVITY CONTROLLER

INSTRUCTION MANUAL

1.0

GENERAL INFORMATION

The AquaMetrix Model 2200C conductivity analyzer is a versatile industrial microprocessor based
instrument. Setpoints and outputs are programmed through the menu with push buttons on the face
of the instrument. Calibration is achieved from the front panel menu. The instrument may be used
in conjunction with any AquaMetrix MS series cell.

The instrument provides non-isolated 0-5 Vdc, 0-1 mA and isolated 4-20 mA analog outputs. The
analog output may be programmed to represent any segment of the measuring scale. One control
relay, with fully adjustable deadband, and one high-low alarm relay are provided. The control relay
can be programmed to activate in response to rising or falling conductivity.

The integrity of the system is ensured with a watch-dog timer and system alarm. A password
feature protects the stored values.

The instrument is housed in a NEMA 4X enclosure. The standard unit is provided with mounting
hardware for surface mount applications. Both panel and pipe mount kits can be ordered
separately.