13

COMPLETE SETUP PROCEDURE

The setup procedure is used to select the tempera-
ture scale, resolution, and thermocouple type.

NOTE

Selected settings are stored in memory and will
remain in memory even after power is turned off,
or while batteries are being replaced.

SELECTING TEMPERATURE SCALE

Select °C or °F by pressing the °C/ °F key. Each
time the key is pressed the temperature scale will
switch. Switching between °C and °F can be done
at any time during operation.

Each time you turn the instrument on, it will power
up with the same settings that were set when the
unit was last turned off.

SELECTING RESOLUTION

Select 0.1° or 1° resolution by pressing the
0.1°/1° key. This key toggles between the two reso-
lutions. When measuring temperature above 999.9°
or below

150°C the thermometer automatically

autoranges to 1° resolution.

SELECTING THERMOCOUPLE TYPE

Select the thermocouple type, either J, T, E, K, by
pressing the J-T-E-K key. Each push of the key
will step to the next type thermocouple. The se-
lected thermocouple type is indicated by the an-
nunciators in the lower right corner of the display.

Continue with the OPERATING PROCEDURES
section or turn the instrument OFF by pressing the
ON/OFF key again. The display will blank.

Advertising