Pyromation Dual J-T-E-K User Manual

Page 26

Advertising
background image

24

3. Allow the reading to stabilize. If the displayed

temperature is higher or lower than the reference
temperature, use the MAX

s

key to increase

the displayed reading or the MIN

t

key to de-

crease the displayed reading until the reference
temperature is displayed. The MIN

t

or MAX

s

key must be pressed at least once. The CAL
annunciator should be blinking during this
procedure.

4. Press the HOLD

key to lock the offset cali-

bration in and advance to the slope calibration.
To return to normal operation press any key ex-
cept CAL or HOLD

.

5. Slope Calibration: Place the probe at the higher

reference temperature.

6. Allow the reading to stabilize. If the displayed

temperature is higher or lower than the reference
temperature, use the MAX

s

key to increase

the displayed reading or the MIN

t

key to de-

crease the displayed reading until the reference
temperature is displayed. The MIN

t

or MAX

s

key must be pressed at least once. The CAL
annunciator should be blinking during this pro-
cedure.

7. Press the HOLD

key to lock the calibration

in.