Ar ch ive d – Seametrics DL75 v.1 User Manual

Page 4

Advertising
background image

+

S

-

DL75

To Flow

Sensor

BA

TTER

Y

Sensor Input

Outputs

Sensor

Pulse

-

+

- +

To Pulse Responsive

Metering Pump

FT415

BA

TTER

Y

LT-11568-D

12/18/06

SeaMetrics incorporated • 19026 72nd Avenue South • Kent, Washington 98032 • uSA

(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • www.seametrics.com

TROuBLEShOOTinG

Problem

Probable cause

Try...

Laptop screen reads
”No connection,” but
“Active” light is NOT
blinking

Laptop screen reads
”No connection,” but
“Active” light IS blinking

Flow rate incorrect

No flow rate

DL75 is transmitting data.
Wait two minutes before taking action.

Dead battery in DL75

Loose cable

Bad cable

K-factor setting is incorrect

Damaged flow sensor

Broken wire

Wait for “Active” light to blink

Replace battery

Tighten cable

Replace cable

Make sure K-factor in DL75 matches flow
sensor

Mechanical: See if rotor spins freely
EX Magmeter: See if “Status LED” indicator
is blinking

Check for flow rate

cOnnEcTiOnS

Pulse

Passthru

Pulse

Scaled

Sensor

Input

Power

4-20 mA

To Flow Sensor

S

12-32 Vdc

Loop Power Supply

4-20 mA Device

-

+

+

-

+

-

+

-

+

-

FT420

+

S

-

DL75

BA

TTER

Y

+

-

caution: Do not apply external
power to the DL75.

DL75 TO FT420

DL75 TO FT415

AR

CH

IVE

D

(2/01/08)

Advertising