TABle 4: Flow Rates converted from Feet/sec to Gallons/Minute at various velocities: schedule 40 pipe

Nominal pipe size

Feet / sec

3"

4"

5"

6"

8"

10"

12"

16"

24"

36"

38"

48"

4.6

7.9

12.5

18

31.2

49.1

70

125

259

600

670

1090

11.5

19.8

31.2

45

78

123

174

275

627

1460

1770

2820

23

39.7

62.4

90

156

246

349

551

1250

2910

3530

5640

46.1

79.4

125

180

312

492

698

1100

2510

5830

7070 11280

115

198

312

450

780

1230

1740

2750

6270 14570 17670 28200

230

397

624

900

1560

2460

3490

5510 12530 29140 35350 56400

461

794

1250

1800

3120

4920

6980 11020 25060 58270 70700 112800

691

1190

1870

2700

4680

7370 10470 16520 37600 87410 106050 170000

flow Range.

These sensors are designed to operate at flow

velocities of 0.5 to 30 feet per second (see Table 4, below).
If erratic readings are encountered at low flows, check the

chart to see if flow is below minimum for the pipe size. The
standard shaft and bearings should have a long life at continu-
ous high flow.