Techne TC-5000 User Manual

Page 32

Advertising
background image

31

Parameter

2 Step PCR

3 Step PCR

RT PCR

Heated Lid

105°C

105°C

105°C

Heated lid before program

On

On

On

Pause before program

Off

Off

Off

Initial denaturation

05m00s 94.0°C

05m00s 94.0°C

05m00s 94.0°C

Hot start

Disabled

Disabled

Disabled

Final extension

05m00s 72°C

05m00s 72°C

07m00s 72°C

Final hold

10°C

10°C

10°C

The thermal cycling conditions are different for each of these three templates:

2 Step PCR

3 Step PCR

RT PCR

Cycles

1

Segment

40m00s 45.0°C

Segment

05m00s 94.0°C

Cycles

30

30

40

Segment

00m30s 92.0°C

00m30s 92.0°C

01m00s 94.0°C

Segment

01m00s 68.0°C

00m30s 55.0°C

00m50s 55.0°C

Segment

00m30s 72.0°C

01m00s 72.0°C

The RT PCR template is for those experiments where the starting material is RNA rather than DNA.
RNA requires reverse transcription by a reverse transcriptase enzyme into complementary DNA (cDNA)
prior to the amplification by the DNA polymerase in PCR. The extra 45°C step is for the reverse
transcription and the 94°C step is to denature the reverse transcriptase before the PCR.

WRITING AND EDITING PROGRAMS

There are three options available for creating programs on the TC-5000:

1. Copy and edit one of the pre-written templates.
2. Edit a pre-written template and save it using a different name.
3. Create a new program with user-specific parameters.

There are 5 pre-written templates. These templates cannot be altered; however, they can be copied,
given a new name and then edited to save time when writing a new program.

The Ice Bucket and Ligation 15 Degrees programs are simple temperature holds, with the former
having an infinite hold at 4°C and the latter at 15°C. In both of these programs the heated lid is
switched off and the initial denaturation, hot start and final extension functions disabled.

The 2 Step and 3 Step PCR templates along with the RT PCR template have the following default
parameters: