Temperature display

Select ºC or ºF.

Display Invalid Data

Entered Warnings

When set to ON, shows prompts when programming to warn the user if

a parameter entered is invalid.

Check for New Versions

When set to ON, a message box will be displayed when a new version

of Techne Workbench is available to download.

Program Defaults

Program defaults can be used to set up the preferred default settings for all newly created programs.

Individual programs can be edited as required.

Heated Lid

Use this to set the default heated lid temperature (in general

approximately 10ºC higher than the hottest block temperature in a

program).

Preheat Lid

Set to ON to pre-heat the lid to the set temperature before the thermal

cycling program begins.

Sample Cooling

Set to ON to cool the block to 4ºC while the lid pre-heats.

Pause

Set to ON to pause at the ambient temperature of the block before the

program starts (unless Sample Cooling has been set to ON; in this
case the unit will pause with the block held at 4ºC). The program will not

commence until the unit is manually un-paused.

Initial denaturation

Set to ON for the required hold time to denature the sample and/or

activate the enzyme before the thermal cycling program begins.