Thermo Fisher Scientific CyberScan DO 300 User Manual

Page 38

Advertising
background image

Instruction Manual

DO 300

35

8.10

rSt: Resetting to Factory Default Setting

This program lets you reset all parameters to factory default settings. This clears all calibration

data, memory, and any other setup functions you might have changed. However, the clock

settings will remain as you have set them.

There are also some other parameters that retain settings when reset is done. See Factory

Default Setting on page 51 for a table of factory default condition.

From measurement mode

1.

Press SETUP key to enter SETUP mode.

2. Press

MI/

or MR/

keys to scroll through main

groups until you view parameter “rSt” in the upper

display.

3.

Press the ENTER key.

4. Press

MI/

or MR/

keys to toggle between NO

and YES.

NO retains current set up parameters.

• YES resets parameters to factory default

condition. (Refer to Section 16 - Factory

Default Setting )

5.

Press the ENTER key to confirm selection and to

return to measurement mode.

Figure 33: Reset to factory

defaults