35
8.10
rSt: Resetting to Factory Default Setting
This program lets you reset all parameters to factory default settings. This clears all calibration
data, memory, and any other setup functions you might have changed. However, the clock
settings will remain as you have set them.
There are also some other parameters that retain settings when reset is done. See Factory
Default Setting on page 51 for a table of factory default condition.
From measurement mode
1.
Press SETUP key to enter SETUP mode.
2. Press
MI/
or MR/
keys to scroll through main
groups until you view parameter “rSt” in the upper
display.
3.
Press the ENTER key.
4. Press
MI/
or MR/
keys to toggle between NO
and YES.
•
NO retains current set up parameters.
• YES resets parameters to factory default
condition. (Refer to Section 16 - Factory
Default Setting )
5.
Press the ENTER key to confirm selection and to
return to measurement mode.
Figure 33: Reset to factory
defaults