Thermo Fisher Scientific CyberScan CON 1500 User Manual

Page 26

Advertising
background image

Instruction Manual CyberScan CON 1500

22

save

view

3. Press the enter key to go into the Set the Single/Multi Point

Calibration Mode (SPCAL) setup page.


4. Use the ▲/save or ▼/view keys to select the options of YES

(single-point) or NO (multi-point) so as to configure the
Calibration Mode to Single-point Calibration or Multi-point
Calibration mode.























5. Press the enter key to confirm selection OR press std to exit

from this page.


6. Continue to access other setup pages using ▲/save or ▼/view

keys OR press std key to exit from the setup and return to
measurement mode.


.