Thermo Fisher Scientific CyberScan CON 1500 User Manual

Page 50

Advertising
background image

Instruction Manual CyberScan CON 1500

46

5. After adjusting to the std solution value (which will be displayed at

the upper display), press enter to initiate the standardisation and
on successful completion, meter returns to measurement mode

OR

Press std to exit from the standardisation page.


)

Adjustable standardisation window provided is ±40% of current
reading.

If the standardisation is successful, then meter displays buffer icon in
the middle part of the display in the measurement mode.

There are separate standardisation provided for Conductivity,
Resistivity and TDS.