- 7 -

2.3.3

Connecting pH/ORP Electrode

1. If the pH/ORP electrode has a BNC connector, remove the BNC

connector from the cable.

NOTE: Oakton Instruments offers an optional ‘BNC to Spade Lug
adapter’ (Order code: 05994-90) that can be used with pH/ORP
electrode without removing the BNC connector.

2. Strip the insulation of the cable so that the bare wires are exposed

enough for connection as shown in Figure 3.

Figure 3 : pH/ORP Electrode Cable

NOTE: Make sure to strip inner black layer (screen) to expose the
clear sheath.

3. Connect pH sensing wire to Pin 1 of J204 connector

4. Connect pH reference wire to Pin 2 of J204 connector










Insulation

pH Referenc

Clear sheath

pH Sensing W

Screen

e Wire

ire