8 troubleshooting – Thermo Fisher Scientific Alpha pH 600 RF (In the process of updating the brand guideline & product image) User Manual

Page 41

Advertising
background image

- 35 -

8

T R O U BL ESHO OTI N G

Cause

Problem

Solution

Power on, but no display a) Loose connections

b) Incorrect

output

voltage

of the power adaptor

a) Ensure cables make good

contact

b) Use an power adaptor with

specified output voltage

Unstable pH reading

a) Dirty electrode
b) Electrical

noise

interference

a) Clean

electrode

and

recalibrate

b) Switch to Symmetric mode

Oscillating temperature
readings

a) Electrical

noise

interference


a) Ensure shield wire is

properly connected to pin 7

Slow response

a) Dirty / Oily electrode

a) Clean electrode

Blinking ATC

a) No temperature probe

connection during ATC
mode

a) Ensure temperature sensing

cable makes good contact

Blinking
electrode
annunciator

a) Error in calibration

a) Ensure calibration standard

solution is not
contaminated. Ensure
electrode is clean.

Or (pH)

a) pH electrode is not

connected

a) Ensure electrode makes

good contact with
Transmitter

Or (Temperature)

a) Temperature probe is

not connected when
ATC enabled

a) Ensure electrode makes

good contact with
Transmitter