5.5 Setting Temperature Coefficient


5.5.1 Temperature Coefficient
This menu allows you to set the Temperature Coefficient
corresponding to the solution whose Conductivity is being
measured. In most controllers, this is fixed at 2.10% per

o

C.

However, for alpha CON 100 C/CX models, it is adjustable
from 0.0 to 10.0% per

o

C (please refer to Appendix 3 for

Temperature Coefficient determination and limits).

1. Enter this menu by pressing

the MODE key to the “SEt”
display if you are in the
measurement mode.

2. Press ENTER key to access

the SETUP menu. Press

key twice to enter the “SEt tc
menu. If you do not wish to
effect any change and keep
the Temperature
Compensation as the previous
value, press MODE key. This
will take you out of this mode
and back to its measurement
mode.

3. Press

ENTER and the display

shows the current value of

Temperature Coefficient (blinking).

4. Use

or

key to change the value of the

Temperature Coefficient. Press ENTER to confirm the
value to be used. Press MODE key to exit to
measurement mode.

SETUP

MODE

ENTER

tc

SETUP

MODE

ENTER

2.10

SETUP

MODE

ENTER

SEt