Momentary ER 70 -

1. May occur when incrementing or decrementing a

Controller unable to respond

value. Error clears itself.

within 250 milliseconds

Momentary ER 71 -

1. Computer program must wait longer for a response

Byte received before the

from the instrument.

response was transmitted

Momentary Er 72 -

1. Data received not valid, possible corruption on the

Incorrect Block Check

comm line. Possible noise problems. Check baud

character was received

rates, instrument addresses, line connections, and
termination.

Momentary Er 73 -

1. Check baud rate.

Byte received with incorrect

2. Possible noise problems. Check instrument

parity.

addresses, line connections, and termination.