56 appendix d - specifications – West Control Solutions MRC 7000 Recorder Manual User Manual

Page 56

Advertising
background image

56

Appendix D - Specifications

Measurement Error Limit

• Type J, K, T, E, N, C T/C's and RTD +/- 0.25% of
reading plus 1 degree @ 25 degrees C

• Type R, S, B T/C's +/- 0.25% of span @ 25 degrees C

• mA, mV, and VDC +/- 0.25% of scaled span plus 1
Whole Digit @ 25 degrees C

Ambient Temperature Error

0.01% of span per degree C deviation from 25 derees C

Scan Rate

1 scan/second

Display Decimal Positions

One, two or three decimal places (0.1 or 1 degree for
T/C or RTD)

Noise Rejection

Normal mode, 85dB minimum at 60Hz or greater.
Common mode, 90dB minimum, 115VAC maximum.

Line Voltage

115/230VAC +/- 1-% 50/60 Hz

Power Consumption

25VA maximum

Operating Temperature

32 degrees to 131 degrees F
0 degrees to 55 degrees C (ambient)

Storage Temperature

-40 degrees to 149 degrees F
-40 degrees to 65 degrees C

Humidity

0 to 90% RH, noncondensing

Dimensions

13.19 H x 15.13 W x 3.63 Deep

Weight

20 pounds maximum

Sensor Fault Detection

Displays SnSr for sensor or transmitter break. Alarms go
off. Relays are de-energized.

Fault detection is not

functional for 0-5V or 0-20 mA inputs.

Agency Approvals

UL Spec. 1092 File#E67237
CSA Spec. C22-2 File#LR39885

Transmitter Power Supply

Provides up to 40mA of current at 24 VDC

Warranty

3 years, see back page for details