A.9

PERFORMANCE

Reference Conditions

Ambient Temperature:

20

°

C

±

2

°

C

Relative Humidity:

60 - 70%

Supply Voltage:

90 - 264V AC 50Hz

Source Resistance:

<10

for thermocouple input

Lead Resistance:

<0.1

/lead balanced (Pt100)

Performance Under Reference Conditions

Common Mode Rejection:

>120dB at 50/60Hz giving negligible effect

at up to 264V 50/60Hz.

Series Mode Rejection:

>500% of span (at 50/60Hz) causes

negligible effect.

Thermocouple Inputs

Measurement Accuracy:

±

0.05% typical,

±

0.1% maximum, of span

±

1LSD. NOTE: Reduced performance with

Type “B” Thermocouple below 600

°

C

(1112

°

F).

Linearisation Accuracy:

High resolution ranges - better than

±

0.5

°

C

Low resolution ranges - better than

±

0.05

°

C

Cold Junction

Compensation:

±

0.3% typical,

±

0.5% maximum.

RTD Inputs

Measurement Accuracy:

±

0.02% typical,

±

0.06% maximum, of span

±

1LSD

Linearisation Accuracy:

High resolution ranges - better than

±

0.5

°

C.

Low resolution ranges - better than

±

0.05

°

C.

Operating Conditions

Ambient Temperature:

0

°

C to 55

°

C