Data Logging with DataLoggerXP

19

XP2i Operation Manual

Keypad usage

The following buttons and button combinations are used to control data logging runs on the XP2i. Where two keys are indicated, the operation is executed by
pressing and holding down the first key, then pressing and holding down the second key until the action is completed.

Note:

Start by turning off the gauge. Then use the following keypad shortcuts, in the order they appear.

X

units + Power: setting Logging Parameters (Mode and Interval)

units

+

Used only when the gauge is turned off, this allows you to select the logging type and interval from the keypad.

Press the

(

peak

)

button repeatedly to cycle through the logging types. Press the

(

units

)

button to accept the format.

Actual

Average

Average with Peaks

On Demand

Actual Ultra Low Power

For further information, see

ultra Low Power (uLP) Mode on page 21

X

setting the Logging Interval

The display will advance to the logging interval, displayed in seconds. The

(

peak

)

button lengthens the logging interval. The

(

zero

)

button shortens the

logging interval. The

(

units

)

button accepts the logging interval and the display reverts to normal.

Note:

At this point, the record icon (

) will illuminate on the display.

Note:

If you select the On Demand logging type, you will not be asked to select a logging interval.