Specifications

5

30 Series Operation Manual

Specifications

Note:

Accuracy includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.

Exposure to environmental extremes of temperature, shock, or vibration may warrant a more frequent period.

Pressure

0 to Full Scale . . . . . . . . . . . . . . . . . . . . ±(0.05% of Reading + Floor Term

*

)

*

Floor Term = 0.005% of FS or display resolution, whichever is greater.

See

Pressure Ranges, Display Scales, & Resolution

on page 8.

Vacuum

±(0.25% of Reading + Vacuum Floor Term

)

Vacuum Floor Term (VFT):

For 16 & 36 PSI: VFT = 0.004 PSI

For 300 PSI: VFT = 0.01 PSI

For 1 or 2 bar, & 100 or 200 kPa: VFT = 0.0003 bar or 0.03 kPa

For 20 bar or 2000 kPa: VFT = 0.001 bar or 0.1 kPa

Vacuum operation is not specified for 600 PSI (40 bar/4000 kPa) models and higher. However, all models can be safely connected to vacuum.

The 30 Series calibrator is not recommended for continuous use at high vacuum. Refer to

XP2i-DP

data sheet for gauges that are intended for continuous

high vacuum use.

Temperature

Operating . . . . . . . . . . . . . . . . . . . . . . . 0° to 50°C (32° to 122°F)

Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification.

Storage. . . . . . . . . . . . . . . . . . . . . . . . . . -20° to 70°C (-4° to 158°F)

Battery should be removed if stored for more than one month.

SI (metric) Versions

“SI” is the French acronym for the International System of Units. 30 Series products with “KPA” in the part number are intended for markets where only SI units

are permitted. Therefore, these models only have kPa and/or MPa, bar and/or mbar available.