Specifications

6

30 Series Operation Manual

Media Compatibility (LP and HP)

Liquids and gases compatible with PTFE (Polytetrafluoroethylene) penetrated, hard anodized aluminum (transducer housing), 316 stainless steel (sensor), and
Viton

®

(O-ring). (Viton is a registered trademark of Dupont Dow Elastomers.)

Products with the letters “SPSS” in the suffix of the part number (e.g., “IS33-36/3000-SPSS”), have stainless steel manifolds and therefore the only wetted mate-
rials are stainless steel and Viton.

Pressure Conversions

1 PSI =

27.6806 inches of water column (water at 4°C [39.2°F])

703.087 millimeters of water column (water at 4°C [39.2°F])

70.3087 centimeters of water column (water at 4°C [39.2°F])

2.03602 inches of mercury (mercury at 0°C [32°F])

51.7149 millimeters of mercury (mercury at 0°C [32°F])

6.8948 kilopascals

0.070307 kilograms per square centimeter

0.068948 bar

68.948 millibar

0.0068948 megapascals

Note:

Other conversions may have been specified at time of order. Refer to your certificate of calibration for details.

Electrical

Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 55 mA

Resolution . . . . . . . . . . . . . . . . . . . . . . . 0.001 mA

Accuracy . . . . . . . . . . . . . . . . . . . . . . . . ±(0.025% of reading + 0.001 mA)

Maximum Voltage . . . . . . . . . . . . . . . 30 VDC

Maximum Current . . . . . . . . . . . . . . . 100 mA

Note:

When connecting to circuits that included HART transmitters, a 250 ohm resistor must be placed in series with the calibrator, to prevent damage to the

calibrator and any other equipment connected to the loop. mA can be displayed as a percentage, where 0 to 100% corresponds to either 4 to 20 mA or

10 to 50 mA.