Specifications – Crystal XP2i-DP Digital Differential Pressure Gauge User Manual

Page 4

Advertising
background image

Specifications

2

XP2i-DP Operation Manual

Specifications

differential pressure accuracy

20 to 100% of Full Scale . . . . . . . . . . ± (0.1% of Reading + Static Line Effect).

0 to 20% of Full Scale . . . . . . . . . . . . ± (0.02% of Full Scale + Static Line Effect).

Static Line Effect: 0.0003 psi/psi of lowest static pressure.

Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.

temperature

Operating . . . . . . . . . . . . . . . . . . . . . . . -10 to 50° C (14 to 122° F).

Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification.

Storage . . . . . . . . . . . . . . . . . . . . . . . . . . -40 to 75° C (-40 to 167° F).

Battery should be removed if stored for more than one month.

Media compatibility

Liquids and gases compatible with 300 Series stainless steel and buna-n (o-ring).

pressure conversions

1 psi =

27.6806 inches of water column (water at 4° C [39.2° F])

27.7070 inches of water column (water at 15.6° C [60° F])

27.7292 inches of water column (water at 20° C [68° F])

2.03602 inches of mercury (mercury at 0° C [32° F])

51.7149 millimeters of mercury (mercury at 0° C [32° F])

703.087 millimeters of water column (water at 4° C [39.2° F])

0.070307 kilograms per square centimeter

68.948 millibar

6.8948 kilopascals

0.068948 bar

0.006895 megapascals