4

XP2i-DP Operation Manual

approved Batteries

The XP2i-DP is Intrinsically Safe only if powered by one of the following battery types:

approved Battery type

ta=

Marking

Rayovac Max Plus 815

-20 to 50° C

Ex ia IIB T4 Ga

Energizer E91

*

Energizer EN91

*

-20 to 50° C

Ex ia IIB T3 Ga

Duracell MN1500

replace batteries with approved type in non-hazardous locations only

*

Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.

Many other battery types and models have been tested but failed to meet the requirements for Intrinsic Safety—do not assume other models are equivalent.
The XP2i-DP can be operated from an external power supply (P/N 2984).

Do not mix battery types or manufacturers.

X

csa: the Xp2i is Intrinsically safe only if powered by one of the following battery types:

approved Battery type

Marking

Rayovac Max Plus 815

Class I, Division 1, Groups A, B, C, D T4

Energizer E91

*

Class I, Division 1, Groups A, B, C, D T3C

Energizer EN91

*

Rayovac Max Plus 815

!

WARNING: use of duracell Mn1500 batteries is not approved for csa certification.

!

WARNING: do not use the ac adapter kit in hazardous locations.

Battery Life . . . . . . . . . . . . . . . . . . . . . . 1500 hours typical (alkaline battery).

Ultra Low Power . . . . . . . . . . . . . . . . . > 1 year, typical with 20 minute recording interval in LT5 mode.

Battery Indicator . . . . . . . . . . . . . . . . . 3-segment Battery Icon:

(

) : Full Battery

(

) : Used Battery

(

) : Low Battery

Dead Battery Indication . . . . . . . . . .