Specifications – Crystal 2100 Series User Manual

Page 4

Advertising
background image

PN: 2102
8/01
Page 4 of 8

Specifications

Battery: 9V alkaline. Eveready #522 or equivalent.

Battery life: 2120 - 300 hours, 2121 - 100 hours.

Low Battery Indication: LCD displays “LOBAT” and/or “LOW BATTERY”

Size: Approximately 86mm x 52mm x 168mm (2

″ x 3½″ x 6½″)

Weight: 340-425g (12-15 ozs.) with battery, depending on model.

Storage Temperature Range: -20 to 70°C (-4 to 158°F)

Operating Temperature Range: 0 to 50°C (32 to 122°F)

Pressure

Pressure Ranges: Refer to the chart on page 5.

Accuracy: ±0.1% F.S.* @ 23°C ±5°C. This includes errors due to

linearity, hysteresis and repeatability. The gauge must be
adjusted to read zero at barometric pressure (except for
absolute pressure versions). *or 1 “least significant digit”
on ranges with limited resolution.

Display: 3½ digit Liquid Crystal Display with ½

″ digit height.

Temperature Stability: ±0.018% F.S. per °C (maximum) from 0 to 18°C and 28 to

50°C.

Compensated Temperature Range: 0 to 50°C (32 to 122°F)

Media Compatibility: Liquids and gases compatible with PTFE (polytetrafluoro-

ethylene) penetrated, hard anodized aluminum transducer
housing, 316 stainless steel (sensor), buna-n (0-ring)

Overpressure rating: 3 times highest pressure range.

Pressure fitting: Dual

1

/

8

″ female NPT (tapered).

Milliamp Meter

Accuracy: ±0.05% of reading (+2 digits on 20 mA scale)@ 23°C

±5°C.

Ranges: 0 to ±19.999 mA and 0 to ±50.00 mA on 50 mA scale.

Display: 4½ digit Liquid Crystal Display with 0.4

″ digit height.

Resolution: 0.001 mA on 20 mA scale, 0.01 mA on 50 mA scale.

Temperature Stability: ±0.01% of reading per °C from 0 to 18°C and 28 to 50°C.

Burden Voltage: 0.21V @20mA: 0.51V @ 50mA.

Overload Protection: Internal 2A/8AG/250V user-replaceable fuse.

Options: Super flexible silicone test lead set - individual Red and

Black banana to banana test leads with hard plastic
insulated banana to alligator clips. Order Part#: AC-1351

Options

1.

High Resolution Pressure Display: 4½ digit Liquid Crystal Display. Specify option HR.
Increases resolution on all ranges by one decimal place.

2.

Pressure fitting kit. Part#: AC-1391. Includes three

1

/

8

″ NPT to ¼″ tube adapters and

two 24

″ pieces of ¼″ polyurethane flexible tubing. For use up to 90 PSI.

3.

Quick disconnect set. Part #: AC-1393. One female coupling body with shutoff to male

1

/

8

″ NPT and one coupling insert to ¼″ plastic tubing adapter.

4.

Low pressure pump. Part #: P-102. Includes a variable volume pressure control and
bleed valve. For up to 60 PSIG.

5.

Low Pressure Plastic Tubing. Part #: AC-1398. Polyurethane tubing, 0.25

″ O.D.,

0.126

″ O.D., 90 PSI working pressure.

Advertising