Calibration procedure – Crystal 2100 Series User Manual

Page 6

Advertising
background image

PN: 2102
8/01
Page 6 of 8

If you have any questions regarding calibration or repair, please contact us at any of the phone
numbers listed on the back of the manual.

Pressure Indicator Calibration Procedure

Use this procedure to calibrate the 2100 Pressure indicator.

1. Verify the condition of the battery. Make sure that the

L0BAT indicator is not on, or check

battery voltage. Battery voltage must be greater than 7V. Replace the battery if neces-
sary.

2. Connect the instrument to the pressure calibration system. Be sure that the pressure

calibration system and the connection to the instrument being calibrated is leak free.

3. Allow the instrument to stabilize at room temperature, away from drafts, for at least 30

minutes before proceeding with calibration.

4. Set the scale to 200

″H

2

O range (or the most sensitive range [selector switch set fully

counter-clockwise]).

5. The front panel

ZERO potentiometer should be centered. To center the front panel ZERO

rotate the front panel

ZERO potentiometer fully clockwise until a click is heard, or there is

no more increase in output. Then rotate the front panel

ZERO pot counter-clockwise 15

turns.

Calibration Procedure

A calibration cycle of 1 year is recommended to maintain the 2100 within specifications. Severe
operating conditions (extreme temperature changes, vibration, etc.) may require a more frequent
calibration cycle.

These procedures assume that you are familiar with proper calibration techniques especially
with regard to high vacuum measurements. (Specifically, a leak free system is required so that
pressure differentials do not occur between the instrument being calibrated and the pressure
standard.) Recommended equipment is listed in Table 1.

Instrument

Minimum Specification

Recommended Model

Pressure Standard

0.025% of scale

Pressurements deadweight
gauge, Model T2300

Precision Current Source

0.005% of reading +1µA

EDC Model 521

Absolute Only:

Vacuum Gauge

±1milliTorr @10 milliTorr

Hastings VT6S2

Absolute Gauge

0.01% of Scale

Ruska PPG-30

Vacuum Pump

capable of 1 milliTorr
dead-end vacuum

Table 1: Required Equipment

Advertising