10. ANALYZER PROBLEM SOLVING


If any problems are not solved with these solutions, contact us or an authorized repair center.

Fault symptom

Causes / Solutions

• Oxygen too high
• CO2 too low

• Air leaking into probe, tubing, water trap, connectors

or internal to analyzer.

• Oxygen cell needs replacing.

• Oxygen reading (- - - -)
• CO reading (- - - -)

• Analyzer was stored in a cold environment and is not

at normal working temperature.

• Oxygen cell or CO sensor needs replacing.

• Pump is switched off.

• Batteries not holding charge
• Analyzer not running on mains

adapter

• Batteries exhausted.
• AC charger not giving correct output.
• Fuse blown in charger plug.

• Analyzer does not respond to flue

gas

• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working or damaged with contaminants.

• Net temperature or Efficiency

calculation incorrect

• Ambient temperature set wrong during Automatic

Calibration.

• Flue temperature readings erratic • Temperature plug reversed in socket.

• Faulty connection or break in cable or plug.

• T flue or T nett displays (- - - -)

• Probe not connected.

• X-Air, EFF, COa or CO2 display
(- - - -)

• Oxygen reading is above 18%.

• Analyzer just continually beeps

• Turn dial back to MENU and press

.

• BAT only shows 65 with fully

charged NiMh batteries fitted

• This is not a problem and is to be expected as NiMh

batteries only deliver 1.25 V per cell whereas
Alkalines deliver 1.5 V per cell. Fresh alkalines might
give a BAT value of 80 or so.