FUTEK IPM52x (D520) Digital Display User Manual

Page 26

Advertising
background image

25



Use SELECT key to choose which function you wish to program.
Then press ENTER key.

"F1 Set" thru "F4 Set" programs the RUN mode
function of the 4 F keys
"r in 1" thru "r in 4" programs the function of the 4
remote inputs
"PEAK 1" programs the peak or valley function of each
channel
"PEAK 2" programs second the peak or valley function
of each channel
"SCALE2" programs the second scaling parameters of
each channel
"CodE" programs the lockout code for authorized entry
into SETUP
"dEFAUL" sets meter setup back to default setup
"rEs" sets the resolution for the meter (4.5 or 5.5 digits)
"A2d" - For factory use only

1. "F1 Set", "F2 Set", "F3 Set" and "F4 Set" - RUN MODE

function key setup

This programs the function of the 4 front panel F keys.
A. "rESEt" assigns F key for resetting latched outputs

• Pushing F key will reset all relays in an alarm

condition.

B. "SCALE2" assigns F key to initiate 2

nd

scaling to be

applied.

• Pushing F key will switch displayed channel to

2

nd

scaling.

• Front panel S2 LED will light to indicate 2

nd

scaling is applied.

• Push F key while S2 is selected will change

scaling back to initial scaling

• If different channel is selected while 2

nd

scaling

is selected, it will also use the 2

nd

scale factor

applied to it.

Advertising